Instrument Inspection & Function Testing – CSSD MCQ Questions and Study Notes
Inspection confirms instruments are clean, intact and working before packaging. Practice free CSSD MCQ questions on this topic below.
Take the Instrument Inspection & Function Testing practice test All 45 chapters
Learning objectives
- Understand the key concepts of this chapter
- Apply them in safe practice
Study notes
Inspect each instrument under good lighting, with magnification when needed, for residual soil, corrosion, cracks, pitting and damaged tips.
Function tests include checking scissors for sharpness, hemostat ratchets for holding, and needle holders for jaw alignment. Lumens are checked for debris. Damaged items are removed and tagged for repair.
Important points
- Inspect after cleaning, before packaging
- Rewash any item with visible soil
- Remove damaged instruments
- Lubricate with steam-permeable instrument milk only if per IFU
Key terms
Function test Ratchet Pitting Magnification Instrument milk
Practice MCQs on Instrument Inspection & Function Testing
Tap “Show answer” to check each answer and read the explanation. These practice questions are written for learning with AI assistance and may contain mistakes. Please report any error so it can be fixed.
Questions 1 to 10
1. Instruments are inspected:
- Never
- After cleaning and before packaging
- Only in the OR
- After sterilization only
Show answer
Inspection follows cleaning.
2. An instrument with visible soil should be:
- Packaged anyway
- Lubricated
- Sterilized twice
- Returned for re-cleaning
Show answer
Soil blocks sterilization.
3. Magnification is used to:
- Label
- Sterilize
- Weigh
- Detect fine debris or damage
Show answer
It shows small defects.
4. Hemostat ratchets are tested to confirm they:
- Rust
- Rotate
- Glow
- Hold securely
Show answer
A weak ratchet slips.
5. Cracked or pitted instruments should be:
- Painted
- Removed from use
- Wrapped anyway
- Ignored
Show answer
Damage harbours soil.
6. Scissors are tested for:
- Color
- Sharpness and alignment
- Length only
- Weight
Show answer
Cutting function.
7. Instrument milk should be:
- Used per IFU and steam-permeable
- Applied to lumens
- Oil-based always
- Never dried
Show answer
Only approved lubricants.
8. Lumens should be checked for:
- Price
- Color
- Residual debris
- Logo
Show answer
Lumens hide soil.
9. Corrosion may indicate:
- Perfect care
- Sterility
- New item
- Poor cleaning, drying or water quality
Show answer
Corrosion has causes to fix.
10. Damaged items are usually:
- Sold
- Tagged for repair or replacement
- Sterilized twice
- Hidden
Show answer
Never return to circulation.
Questions 11 to 20
11. Which method can help verify cleaning by detecting organic residue on devices?
- Weight check
- Bowie-Dick test
- ATP bioluminescence testing
- Steam printout
Show answer
US practice: ATP, protein or hemoglobin tests are used for cleaning verification.
12. Cracked insulation on an electrosurgical instrument risks:
- Rust only
- Faster cutting
- Patient or staff burns
- Better grip
Show answer
US practice: inspect and test insulation.
13. Instruments with stains or discoloration require:
- Extra wrap
- Ignoring
- Evaluation per IFU; remove if damaged
- Painting
Show answer
US practice: stains may indicate residue or damage.
14. Lighted magnification is used to inspect:
- Box locks, serrations, lumens
- Only labels
- Only wraps
- Only handles
Show answer
US practice: hard-to-clean areas hide soil.
15. A sticking box lock should be:
- Left as it is because it still closes
- Soaked in saline overnight
- Forced open with a hammer
- Cleaned/lubricated per IFU or removed for repair
Show answer
US practice: functional testing before packaging.
16. A ratcheted instrument that fails to hold should be:
- Used with extra force
- Washed again
- Packed anyway
- Removed for repair or replacement
Show answer
Function must be reliable.
17. Which scissor defect means the instrument should be removed?
- A clean shine
- A short label
- Dull, misaligned or nicked blades
- Closed rings
Show answer
Blades must cut cleanly.
18. What should be checked on forceps jaws?
- Alignment of the tips and teeth
- The weight
- The brand
- The color
Show answer
Jaws must meet properly.
19. A cracked ring handle means the instrument:
- Needs more lubricant
- Needs repair or replacement
- Can be packed
- Is fine for use
Show answer
Cracks weaken the instrument.
20. Why use a borescope on a lumen?
- To weigh the lumen
- To label it
- To see hidden debris inside narrow channels
- To dry it
Show answer
Visual proof of cleanliness.
Questions 21 to 30
21. Inspection covers:
- Only the box
- Handles, joints, shanks and tips
- Only the weight
- Only the label
Show answer
Check every surface.
22. Defects to look for include:
- Weight only
- Color only
- Pitting, cracks, burrs and bent parts
- Shine only
Show answer
Any damage matters.
23. Hemostat jaw teeth should:
- Be missing
- Be flat
- Be bent
- Mesh fully when closed
Show answer
Needed for grip.
24. On a needle holder you check that the jaw inserts are:
- Painted
- Secure and not loose
- Heavy
- Warm
Show answer
Loose inserts fall out.
25. Scissor sharpness is tested by:
- Tapping them
- Weighing them
- Cutting a test material
- Washing again
Show answer
Use a test fabric.
26. Forceps tips should be:
- Rusty
- Aligned without gaps or bends
- Bent
- Loose
Show answer
Tips must meet.
27. Retractor blades are checked for:
- Weight
- Color
- Cracks, especially at welds
- Smell
Show answer
Weak welds fail.
28. A ratchet is tested by closing to the first tooth and:
- Tapping lightly to check it holds
- Hammering
- Heating
- Washing
Show answer
Simple function test.
29. A box lock should move:
- Not at all
- Loosely
- Smoothly without sticking
- Stiffly
Show answer
Stiff locks need service.
30. Lubricant after inspection is applied:
- With cooking oil
- Lightly, using an approved product
- Never
- Heavily
Show answer
Per the IFU.
Questions 31 to 40
31. An instrument with rust should be:
- Painted
- Dried more
- Packed anyway
- Removed from the set
Show answer
Rust spreads.
32. Stained instruments are:
- Investigated to find the cause
- Packed
- Ignored
- Hidden
Show answer
Stains may be residue.
33. Suction tubes are checked for:
- A clear lumen
- Weight
- Labels
- Color
Show answer
Blocked tubes fail.
34. Rigid scopes are checked for:
- Only color
- Lens clarity and damage
- Only weight
- Only labels
Show answer
Damage harms the view.
35. Broken fibers in a fiberoptic cable appear as:
- Dark dots when held to light
- Colors
- Bright dots
- Nothing
Show answer
Light transmission check.
36. Laparoscopic instruments are checked for:
- Only labels
- Only weight
- Insulation, tips and straight shafts
- Only color
Show answer
Damage causes burns.
37. Powered tools are checked for:
- Weight
- Smell
- Cords, batteries and function
- Color
Show answer
Per the IFU.
38. Missing parts are:
- Hidden
- Reported and replaced
- Packed anyway
- Ignored
Show answer
Incomplete sets delay surgery.
39. Sets are verified against:
- A calendar
- The count sheet
- A menu
- A timetable
Show answer
Ensures completeness.
40. Trays are checked for:
- Damage, holes and rust
- Color
- Weight
- Smell
Show answer
Damaged trays harm drying.
Questions 41 to 50
41. Silicone mats are checked for:
- Smell
- Color
- Tears and wear
- Weight
Show answer
Torn mats trap soil.
42. Rigid container seals and filters are:
- Removed
- Ignored
- Painted
- Inspected before use
Show answer
Barrier integrity.
43. Inspection lighting should be:
- Flashing
- Colored
- Dim
- Bright and shadow-free
Show answer
See small defects.
44. The inspection area is kept:
- Clean and separate from decontamination
- Open
- Merged with decontamination
- Dirty
Show answer
Prevents recontamination.
45. A defective instrument is tagged with:
- Nothing
- What is wrong with it
- A smiley
- A price
Show answer
Helps repair.
46. Repair tracking records:
- Staff names only
- Colors
- Items sent out and returned
- Lunch
Show answer
Prevents lost instruments.
47. Sharpening is done by:
- Visitors
- A qualified technician
- Patients
- Anyone
Show answer
Poor sharpening ruins edges.
48. Items that fail inspection and are soiled go back to:
- The sterilizer
- The OR
- Decontamination
- Storage
Show answer
Re-clean first.
49. Dull scissors can:
- Cut better
- Tear or crush tissue
- Heal tissue
- Sterilize
Show answer
Dull blades are unsafe.
50. Hairline cracks often appear at:
- The wrap
- The label
- Hinges and joints
- The tray
Show answer
Check these spots.
Questions 51 to 52
51. Flaking chrome plating means the instrument should be:
- Dried
- Packed
- Removed from use
- Oiled
Show answer
Flakes can enter the wound.
52. Microsurgical instruments are protected after inspection with:
- Tape
- Paper
- Tip protectors
- Nothing
Show answer
Prevents bent tips.
Related chapters
- Chapter 1: Decontamination & Cleaning
- Chapter 2: Instrument Cleaning & Decontamination Techniques
- Chapter 4: Packaging & Sterile Barrier Systems
- Chapter 5: Sterilization Principles
Continue learning
CSSD interview questions Mock tests MCQ practice CSSD NABH checklist
Frequently asked questions
How many Instrument Inspection & Function Testing MCQs are available?
This page has 52 free MCQs on Instrument Inspection & Function Testing with answers and explanations. More are added over time.
Is Instrument Inspection & Function Testing practice free?
Yes. You can practice without login or payment.
Are these official CSSD exam questions?
No. They are original practice questions written for learning.